1. Review the UIS setup. Make sure that the system resembles the Figure.
2. Power Control Module (PCM) supplies power to both the lamp and the fan.
3. UIS-IRIS can be added between IS-Lamp housing and integration sphere by screwing in place. IRIS’s provide lamp attenuation without changing lamp Color Temperature (K) of the UIS. Make sure the lamp cools down for 3-5 minutes before exchanging IRIS. 4. Spectral calibration data is supplied with each UIS for Radiance (standard) or Irradiance at 2” port hole at the highest power setting. Total Radiance is then included at subsequent lower levels (i.e. with PCM or IRIS’s); however, complete spectral curves for lower power levels must be purchased additionally (item: UIS-CAL). 5. Using the UIS – Allow 5 minutes warm up and do not leave the lamp on unattended or for extended periods of time. Use tissue paper or kimwipes when touching the bulb directly to avoid leaving fingerprints.
|
Uniform Illumination Sphere – Light Source (UIS-LS) – Quick Start
Tags: Application